※ The bulb is Toshiba 0.3 microfocall length, 80KHZ.
10. Shot time in taking photos
Specifications for portable dental x-ray :
a) The rated parameters of x-ray production device
Tube voltage: | 70kv (fixed) |
Tube current: | 1MA (fixed) |
High voltage generating circuit: | High frequency switch |
X-ray control method: | Microprocessor control |
Time setting range: | 0.05-1.2 seconds |
b) Main unit Net Weight: 1.95kg
c) Size: 215*140*160 mm
d) X-ray tubing
Type: | bipolar fixed x-ray tube |
X-ray tube Model: | G-073B-DC |
The size of x-ray tube focal spot: | 0.3*0.3mm |
ANODE heating unit: | 8.5Khu |
Maximum tube voltage: | 75Kv |
Target angle: | 21° |
Total filtration: | 1.6mmAL (Origin filtration: 0.8mmAL, Additional filtration: 1mmAL) |
X-ray irradiation range: | 73mm (roundness) |
e) Useage voltage: battery DC16.8V
f) Charger: Input AC100-240V, Output: DC16.8V
At present, we are dealing on seven categories like dental equipment, dental orthodontic series, dental supplies, dental material, dental decoration and dental accessories etc, amounts to more than six hundreds kind of products, such as NSK/KAVO dental handpiece, woodpecker brand of curing light and apex locator series ect. autoclave, water distiller, ultrasonic Scaler, orthodontic arch wire and bracket, dental study model, dental chair etc.